Product Description
E-Module
Emergency Module
Size and Weight
- Size: 160mm x 99mm x 71mm
- Weight: < 0.6kg
Display
- 3.5 Color TFT-LCD, Resolution: 320×240 pixels
- Waveform: Up to 12 tracks
Battery
- Type: 3.7V/1800mAh Rechargeable lithium ion battery
- Operating time: >60 minutes
- Charging time: Standby state: <6h
- Turn off delay: 5 to 10 minutes after the low battery alarm first occurs.
- Standard and Interface: HR, SpO2, NIBP, RESP, TEMP, and waveform of ECG, SpO2, RESP.
- (ECG waveform available for concatenation)
- Big font Interface: HR, SpO2, NIBP, 1 channel ECG waveform.
- User defined interface: 3 basic parameter + 1 Blood Pressure parameter, 1-3 channel waveform
- Arrhythmia analysis and ST analysis.
- Simultaneous display of 7/12 lead ECG in one screen
- Independent physiological alarm light and technical alarm light.
- Seamless transport solution, Fully meet the clinic demand.
- Providing fast and convenient monitoring during the patient transfer with display, battery, alarm and storage capability.
- Patient’s data transferring automatically when connect to main Unit.
- Extremely compact design, convenient to carry, allowing patient to roam freely.
- 3.5 Color TFT-LCD display.
- 480 groups NIBP review.
- Power off storage of 8 hours trend.
- Built-in rechargeable lithium battery can work more than 1 hour.
E-Module options:
|
M-1 |
M-2 |
M-3 |
M-4 |
M-5 |
M-6 |
M-7 |
M-8 |
RESP |
√ |
√ |
√ |
√ |
√ |
√ |
√ |
√ |
2-TEMP |
√ |
√ |
√ |
√ |
√ |
√ |
√ |
√ |
NIBP |
√ |
√ |
√ |
√ |
√ |
√ |
√ |
√ |
2-IBP |
√ |
√ |
|
|
√ |
√ |
|
|
3/5 Lead ECG |
√ |
|
√ |
|
√ |
|
√ |
|
12-Lead ECG |
|
√ |
√ |
√ |
|
√ |
|
√ |
Digital SpO2 |
√ |
√ |
√ |
√ |
|
|
|
|
- Touch Screen (Option), supporting for Mouse Key board operation.
- Available for multi-language interface.
- Independent Physiological & Technical Alarm.
- Built-in detachable rechargeable lithium battery
- Particularly backlight button for night operation
- SD card socket, effective storage for historical patent data.
Drug dose calculation
- Hemodynamic calculation
OXYCRG
- Display 16 minutes trend of HR, SPO2, RESP
Short trend
- Maximum 168 hour graphic and tabular trends
Other bed
- Display other bedside monitor’s all parameter and one waveform, support user defined parameter display
Big font
- User can select 5 parameters (related waveform will display of available)
- Clear observation in a long distance, especially suitable for ICU, CCU room checking and monitoring during night
Technical data MACS 50 |
Sternmed Multi Parameters Patient Monitor |
|||||||||||
Size and Weight |
|
|||||||||||
Size |
43mmx89x206mm |
|||||||||||
Weight |
<11kg )A8) |
|||||||||||
Standard module slot |
4 |
|||||||||||
Additional module rack Slot |
1 |
|||||||||||
Power supply |
|
|||||||||||
Power voltage |
AC 100-240V 50/60Hz |
|||||||||||
Power Input |
<150VA |
|||||||||||
Input Current |
1.7-0.8A |
|||||||||||
Safety class |
Category 1 |
|||||||||||
Display |
|
|||||||||||
Resolution |
17’’ color Anti-glare TFT-LCD 1280×1024 pixels |
|||||||||||
Battery (Option) |
|
|||||||||||
Type |
Rechargeable Lithium battery, 11.1V/4.0AH Operating time under the normal use and full charge :>60minutes (2 batteries for 120 minutes) |
|||||||||||
Recorder (Option) |
|
|||||||||||
Method |
Thermal dot array |
|||||||||||
Paper width |
50 mm (1.97 in) |
|||||||||||
Paper length |
15m |
|||||||||||
Paper Speed |
12.5 / 25 / 50 (mm/sec) |
|||||||||||
Traces |
Maximum 3 tracks |
|||||||||||
Recording way |
Real-time recording, Periodic recording, Alarm recording |
|||||||||||
Alarm |
|
|||||||||||
Level |
Low, medium and high |
|||||||||||
Indication |
Auditory and visual |
|||||||||||
Patient Physiological Alarm Light color |
Yellow & Red |
|||||||||||
Equipment: Technical Alarm Light color |
Blue |
|||||||||||
|
Supports Pitch tone and multi-level volume |
|||||||||||
|
Supports custom arrhythmia |
|||||||||||
Input device |
|
|||||||||||
Touch button |
standard configuration, support operation of shortcut menu |
|||||||||||
Knob |
Standard configuration |
|||||||||||
Mouse Input |
Support |
|||||||||||
Keyboard Input |
Support |
|||||||||||
System Output & Extensible Interfaces |
|
|||||||||||
Ethernet Network |
2 Standard RJ45 socket |
|||||||||||
Defibrillation Output |
1RJ11 socket |
|||||||||||
Nurse Call |
1 BNC socket |
|||||||||||
Video Output |
1 DVI port, 1 VGA port |
|||||||||||
USB 1.1 port |
6 |
|||||||||||
Auxiliary Module Rack connector |
1 |
|||||||||||
SD memory card |
2 KG (Standard configuration ) |
|||||||||||
Analog Output (ECG or IBP) |
Option |
|||||||||||
Trend & Reviewing |
|
|||||||||||
Trend |
168 hours |
|||||||||||
NIBP measurement reviewing |
1000 groups |
|||||||||||
ARR event |
128 groups of ARR event and the associated parameter |
|||||||||||
waveform |
at the alarm moment |
|||||||||||
Holographic waveform |
The storage time depends on the stored waveforms and the quantity of them. |
|||||||||||
|
||||||||||||
Technical data MACS 50 |
Sternmed Multi Parameters Patient Monitor |
|||||||||||
Environment |
|
|||||||||||
Operating temperature |
0-+40°C |
|||||||||||
Storage temperature |
-20°C to +50°C |
|||||||||||
Operating humidity |
15% to 85% (non condensing) |
|||||||||||
Storage humidity |
10% to 93% (non condensing) |
|||||||||||
Operating atmospheric pressure |
860hPa to 1060hPa |
|||||||||||
Storage atmospheric pressure |
500hPa to 1060hPa |
|||||||||||
Safety |
|
|||||||||||
|
IEC60601-1 Approved, CE marking according to MDD93/42/EEC |
|||||||||||
Performance |
|
|||||||||||
ECG
|
Lead Mode |
3-leads ECG input 5-leads ECG input 12-leads ECG input |
||||||||||
Lead selection |
I, II, III I, II, III, aVR, aVL, aVF, V- I, II, III aVR, aVL, aVF, V1-V6 (option) |
|||||||||||
Gain |
2.5 mm/mV(x0.25), 5 mm/mV (+0.5), 10 mm/mV(x1), 2.0 mm/mV(x2), 40 mm.mV(x4), Auto |
|||||||||||
CMRR |
Monitor mode >105dB Surgery mode >105dB Diagnostic mode 0.05-150Hz |
|||||||||||
Input impedance |
>5.0 Mohm |
|||||||||||
ECG signal range |
+10.0 mV |
|||||||||||
Patient Leakage Current |
<10 uA |
|||||||||||
Standardizing signal |
1mV+5% |
|||||||||||
Baseline recovery |
5s after Defibrillation. (Mon or Surg mode) |
|||||||||||
Indication of electrode separation |
Every electrode (exclusive of RL) |
|||||||||||
Protection |
Breakdown Voltage 4000VAC 50/60HZ; defibrillator proof |
|||||||||||
Sweep speed |
12.5mm/s, 25mm/s, 50mm/s |
|||||||||||
HR
|
Range |
Adult 10-300 bpm Pediatric & Neonate: 10-350 bpm |
||||||||||
Refreshing time |
<50 bpm Per2 pulses 50-120bpm Per 4 pulses > 120bpm Per 6 pulses |
|||||||||||
Resolution |
1 bpm |
|||||||||||
Accuracy |
+1% or + bpm, whichever is greater |
|||||||||||
ST segment |
|
|||||||||||
Measurement range |
-2.0mV-2.0mV |
|||||||||||
Accuracy |
-0.8V-0.8mV: +1 bpm, whichever is greater Over + 0.8mV: unspecified |
|||||||||||
Resolution |
0.01mV |
|||||||||||
RESP |
|
|||||||||||
Method |
Thoracic impedance |
|||||||||||
Lead Selected from |
I (RA-LA) or II (RA-LL): Default: I |
|||||||||||
Gain |
x0.25,x1, x2, x4 |
|||||||||||
Bandwidth |
0.25 Hz to 2.0Hz (-3db) |
|||||||||||
Sweep speed |
6.25mm/s, 12.5mm/s, 25mm/s |
|||||||||||
Measurement Range |
0-150 rpm |
|||||||||||
Resolution |
1rpm |
|||||||||||
Accuracy |
+2 rpm or 2% whichever is greater |
|||||||||||
Delay of Apnea Alarm |
10s, 15s, 25s,30s, 35s, 40s, 45s, 50s, 55s, 60s |
|||||||||||
|
||||||||||||
Technical data MACS 50 |
Sternmed Multi Parameters Patient Monitor |
|||||||||||
NIBP |
|
|||||||||||
Way of measurement |
Automatic oscillometry |
|||||||||||
Range of measurement |
Adult |
SYS 30-270 mmHg DIA 10-220 mmHg MAP 20-235 mmHg |
||||||||||
Child |
SYS 30-235 mmHg DIA 10-220 mmHg MAP 20-225 mmHg |
|||||||||||
Neonate |
SYS 30-135 mmHg DIA 10-100 mmHg MAP 20-125 mmHg |
|||||||||||
Cuff pressure range |
0-300 mmHg |
|||||||||||
Resolution |
1 mmHg |
|||||||||||
Pressure Accuracy |
Static |
+% or + 3 mmHg, whichever is greater |
||||||||||
Clinical |
+5 mmHg average error |
|||||||||||
Standard deviation |
< 8 mmHg |
|||||||||||
Unit |
mmHg, kPa |
|||||||||||
Measurement mode |
Manual, Auto, STAT |
|||||||||||
Intervals for AUTO measurement time |
1,2,3,4,5,10,15,30,60,90 minutes; 2,4,8,12hours |
|||||||||||
STAT mode cycle time |
Keep 5 minutes, at 5 seconds Interval |
|||||||||||
Overpressure Protection |
Hardware and software double protections |
|||||||||||
Pulse rate range |
40-240 bpm |
|||||||||||
BLT-SpO2 (Digital Technic) |
|
|||||||||||
Measurement Range |
0-100% |
|||||||||||
Resolution |
1% |
|||||||||||
Accuracy |
At 70-100%, +2% At 0-69%, unspecified |
|||||||||||
PR
|
Measurement Range |
25-255 bpm |
||||||||||
Resolution |
1 bpm |
|||||||||||
Accuracy |
+1% or +1 bpm, whichever is greater |
|||||||||||
Nellcor–SpO2 (option) |
|
|||||||||||
Measurement range |
0-100% |
|||||||||||
Resolution |
1% |
|||||||||||
Accuracy |
At 70-100%, +2% (Adult) At 70-100%, +3% (Neonate) At 70-100%, + 2% (Low Perfusion) At 0-69%, unspecified |
|||||||||||
PR |
Measurement Range |
20-300 bpm |
||||||||||
Resolution |
1 bpm |
|||||||||||
|
Accuracy |
20bpm to 250bpm: +3 bpm 251bpm to 300bpm: unspecified |
||||||||||
Masimo SpO2 (option) |
|
|||||||||||
Measurement range |
0% to 100% |
|||||||||||
Resolution |
1 % |
|||||||||||
Accuracy |
70% to 100%: +2 % (adult/pediatric, non-motion conditions) 70% to 100%: +3 % (neonate, non-motion conditions) 70% to 100%: +3 % (motion conditions) |
|||||||||||
Averange time |
2-4s, 4-6s, 8s, 10s, 12s, 14s, 16s |
|||||||||||
PR |
Measurement Range |
25 bpm to 240 bpm |
||||||||||
Accuracy |
+3 bpm (non-motion conditions) +5 bpm (motion conditions) |
|||||||||||
Resolution |
1bpm |
|||||||||||
|
||||||||||||
Technical data MACS 50 |
Sternmed Multi Parameters Patient Monitor |
|||||||||||
TEMP |
|
|||||||||||
Max Channel |
8 |
|||||||||||
Measurement way |
Thermal resistance way |
|||||||||||
Measurement Range |
0.0°C-50.0°C (32°F-122°F) |
|||||||||||
Accuracy |
+0.1°C or 1°F |
|||||||||||
Unit |
Celsius (°C), Fahrenheit (°F) |
|||||||||||
Connecting cable |
Compatible with YSI-400 serial |
|||||||||||
IBP |
|
|||||||||||
Max Channel |
8 |
|||||||||||
Measurement way |
Directly invasive pressure measurement |
|||||||||||
Sensitivity of transducer |
5uV/V/ mmHg, +2% |
|||||||||||
Impedance of transducer |
300 to 3000Ω |
|||||||||||
Measurement Range |
+50-+350mmHg |
|||||||||||
Resolution |
1mmHg |
|||||||||||
Unit |
mmHg, kPa,cmH2O |
|||||||||||
Accuracy |
Static |
+ 1mmHg or +2%, whichever is greater (exclusive of transducer) + 4mmHg or +4%, whichever is greater (inclusion of trasducer) |
||||||||||
Dynamic |
+ 4mmHg or 4 %, whichever is greater |
|||||||||||
Transducer sites |
Arterial Pressure (ART) Pulmonary Artery Pressure (PA) Left Atrium Pressure (LAP) Right Atrium Pressure (RAP) Central Venous Pressure (CVP) Intracranial Pressure (ICP) P1/P2 |
|||||||||||
Selection of measurement range |
ART: 0- +350 mmHg PA: +10- +120 mmHg CVP/RAP/LAP/ICP: -10- +40mmHg P1/P2: +50-+350mmHg |
|||||||||||
EtCO2 (Side-stream) |
|
|||||||||||
Measure method |
Infrared spectrum |
|||||||||||
Measurement Range |
0.0-13.1% (0-99.6 mmHg) |
|||||||||||
Resolution |
1 mmHg |
|||||||||||
Unit |
% mmHg, kPa |
|||||||||||
Accuracy |
0% to 4.9%, +0.3% (+2.0 mmHg) 5.0% to 13.1%, , < +10% of the reading |
|||||||||||
Measurement range of awRR |
3-150 rpm |
|||||||||||
Calibration |
Offset calibration: auto, manual, Gain calibration |
|||||||||||
EtCO2 (Main-stream) |
|
|||||||||||
Measure method |
Infrared spectrum |
|||||||||||
Warm up time |
Capnogram displayed in less than 15 seconds, At an ambient Temperature of 25°C, full specifications within 2 minutes. |
|||||||||||
Measurement Range |
0.0-19.7% (0-150 mmHg |
|||||||||||
Resolution |
1 mmHg |
|||||||||||
Rise time (10 I/min) |
< 60 ms |
|||||||||||
Unit |
%, mmHg, kPa |
|||||||||||
CO2 Accuracy |
0-40 mmHg. + 2 mmHg 41-70 mmHg. +5% of reading 71-100 mmHg, +8% of reading 101-150 mmHg, +10% of reading (at 760 mmHg, ambient temperature of 35°C) |
|||||||||||
awRR measurement range |
0-150 rpm |
|||||||||||
awRR measurement Accuracy |
+1rpm |
|||||||||||
|
||||||||||||
Technical data MACS 50 |
Sternmed Multi Parameters Patient Monitor |
|||||||||||
EtCO2 (Microstream) |
|
|||||||||||
Measure method |
Infrared spectrum |
|||||||||||
Warm up time |
Capnogram displayed in less than 20 seconds, At an ambient temperature of 25°C, full specifications within 2 minutes. |
|||||||||||
Measurement range |
0-19.7% (0-150 mmHg) |
|||||||||||
Resolution |
1 mmHg |
|||||||||||
Unit |
%, mmHg, kPa |
|||||||||||
CO2 Accuracy |
0-40 mmHg, +2 mmHg 41-70 mmHg, +5% of reading 71-100 mmHg, +8% of reading 101-150 mmHg, +10% of reading (at 760 mmHg, ambient temperature of 25°C) (when RR>80rpm, all the range is +12% of reading) |
|||||||||||
CO2 rersponse time |
<3s |
|||||||||||
awRR measurement range |
2-150 bpm |
|||||||||||
awRR measurement Accuracy |
+1rpm |
|||||||||||
Sample Flow Rate |
50ml/mIn + 10ml/mIn |
|||||||||||
Anesthetic Gas |
|
|||||||||||
Measure method |
Infrared spectrum |
|||||||||||
Fi and Et values |
CO2 N2O, O2, AG (HAL, ISO, ENF, SEV, DES) |
|||||||||||
Resolution |
1% |
|||||||||||
Unit |
1% |
|||||||||||
Calibration |
Room air calibration performed automatically when changing airway adapter (<5 sec) |
|||||||||||
Warm-up time |
<10s, full accuracy within 1 min |
|||||||||||
Measurement and alarm range of AG |
Gas |
Range |
Accuracy |
|||||||||
CO2 |
0-10% |
+ (0.3% ABS +4% REL) |
||||||||||
N20 |
0-100% |
+ (2% ABS+8% REL) |
||||||||||
O2 |
10-100% |
+ (2% ABS+2% REL) |
||||||||||
HAL, ISO, ENF |
0-5% |
+ (0.15% ABS+2% REL) |
||||||||||
SEV |
0-8% |
+ (0.15% ABS+10% REL) |
||||||||||
DES |
0-18% |
+ (0.15% ABS+10% REL) |
||||||||||
awRR measurement range |
0-150rpm |
|||||||||||
awRR measurement Accuracy |
+1rpm |
|||||||||||
Rise time (flowing speed 10 I/min) |
CO2 < 90 ms O2 < 300 ms N2O < 300 ms Hal, Iso, Enf, Sev, Des < 300 ms |
|||||||||||
Total system response time |
<1 seconds |
|||||||||||
|
||||||||||||
Technical data MACS 50 |
Sternmed Multi Parameters Patient Monitor |
|||||||||||
Noninvasive Cardio Output (ICG): |
|
|||||||||||
Method |
Measurement of thoracic electrical bio-impedance |
|||||||||||
Measurement Range |
HR: 40-250 bpm SV: 5-250ml SI:5-125mL/m2 C.O.: 1.4-15 L/min TFC: 15-143 KΩ |
|||||||||||
Accuracy |
HR +2bpm SV: unspecified C.O unspecified |
|||||||||||
Alarm range |
C.I.: 0.0 L/min/m2 to 15.0 L/min/m2 continuous I adjustable. TFC: 10 /kΩ continuously adjustable |
|||||||||||
|
|
|||||||||||
Main unit: 17” anti-glare TFT-LCD display, 4 Standard module slot, 1 Additional module rack Slot (for EMS all-in-one module), 13 Touch buttons , 2RJ45 Ethernet socket, 1 Defibrillation Output, 1 Nurse Call Socket, 1 DVI port, 1 VGA port, 6 USB 1.1 port, 1 Auxiliary Module Rack connector,2G SD memory card.
EMS module: 8 kinds of option Option Module: Side-stream CO2 module, Micro-stream CO2 module, Main-stream CO2 module, Main-stream CO2, AG module, ICG module, IBP module, Temp module, SpO2 module Module Rack: Auxiliary Module Rack Navigating: USB compatible mouse and keyboard. Printing: 3 channel thermal recorder Mounting: Rolling stand, wall mount Battery: 11.1V/4.0AH Rechargeable Lithium Battery (max 2 pcs.) Other options: External Display, Wireless LAN, Extensive Memory card, Analog Output (ECG or IBP), Touch Screen.
|
||||||||||||